| Model Number |
SN-665 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when a nurse was pulling out the product, the thread broke.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample two partial sutures.It was observed that the needle was detached from the suture, but the needle was not returned.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failure.If information is provided in the future, a supplemental report will be issued.
|
| |
|