• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L. - CRM FACILITY REPLY; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L. - CRM FACILITY REPLY; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number REPLY DR
Medical Device Problem Code Incorrect Device Or Component Shipped (2962)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Reportedly, on (b)(6) 2018, usa crm technical services was informed that the as-shipped polarities of the subject pacemaker were unipolar instead of bipolar.As reply pacemakers are designed to prevent polarity reprogramming from unipolar to bipolar without normal lead impedance measurement, the polarity cannot be reprogrammed before implantation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPLY
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR  92140
MDR Report Key7519157
Report Number1000165971-2018-00499
Device Sequence Number1385416
Product Code DXY
UDI-Device Identifier08031527011523
UDI-Public(01)08031527011523(11)171115(17)190615
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/15/2019
Device Model NumberREPLY DR
Device Catalogue NumberREPLY DR
Device Lot NumberS0291
Was Device Available for Evaluation? No
Distributor Facility Aware Date04/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2018
Initial Report FDA Received Date05/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-