We have not received the device for evaluation since the device has been implanted in the patient.The physical incubation test (which measures how well the collagen was impregnated into the graft), blood and water permeability previously performed on a different portion of the graft from the same lot number was found to be acceptable.Also, the sister graft which is the another half of the same exact graft had also been implanted successfully and without any reported issues.We have conducted a lot history review and did not find any issues noted in the manufacturing or packaging process that could be related to this issue.All qc tests passed their requirements.We have sold all of the grafts that were manufactured under this lot number.We have not received any other complaints of a similar nature for devices from this lot.Patient has been discharged from the hospital.However, patient is scheduled for follow-up to be examined for any blood clot formation.Protamine sulphate was used to reverse the effects of the heparin.Due to the excessive bleeding, 3 blood transfusions were required to replenish the lost blood.Physician had to use copious amount of glue to reduce the leak.
|