| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip (b)(4).Note: manufacturer contacted customer on 4/27/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for erratic blood glucose test results.The customer is concerned with tests results from back to back blood tests of 61 and 239 mg/dl.Customer was only concerned results were erratic.The customer's expected fasting blood glucose test result range is 100 - 110 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 06/20/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) internal report # 01560483 returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer on 4/27/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for erratic blood glucose test results.The customer is concerned with tests results from back to back blood tests of 61 and 239 mg/dl.Customer was only concerned results were erratic.The customer's expected fasting blood glucose test result range is 100 - 110 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on 04/23/2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 06/20/2020 and open vial date is 04/02/2018.The meter memory was reviewed for previous test result history: result 1 :116mg/dl date:(b)(6) 2018 time:11:12am fasting.Result 2 :87mg/dl date:(b)(6) 2018 time:6:12am fasting.Result 3 :102mg/dl date:(b)(6) 2018 time:7:15am fasting.Result 4 :106mg/dl date:(b)(6) 2018 time:1:17pm fasting.Result 5 :176mg/dl date:(b)(6) 2018 time:10:15am non-fasting.
|
| |
|