MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGES; HEPARIN FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGES; HEPARIN FLUSH Back to Search Results
Catalog Number 306424
Health Effect - Clinical Codes Respiratory Distress (2045); Reaction (2414); Sweating (2444)
Date of Event 04/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This mdr is intended to report 1 of the 2 events described in the event.Another mdr will be filed for the 2nd incident.(1st event submitted with mdr report# 1911916-2018-00229) the customer's address is unknown.Unknown, washington (wa) , 0000 usa has been used as a default.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
 
Event or Problem Description
¿it was reported that on two occasions ((b)(6)), after 30 minutes of pushing iv saline, patient experienced a severe reaction (coughing, couldn't breathe, cold sweats).Patient also used heparin flush but thinks the issue was from the saline more so than the heparin.Patient went to the er on both occasions and was told that nothing was able to be determined.Er checked the patient's lungs, checked for embolism, etc.¿ there was no additional specific information on the medical interventions received.
 
Additional Manufacturer Narrative
Five samples were received by bd for evaluation.A quality engineer was able to inspect the samples and sent them out for ftir testing.The results of the testing were all negative and the samples passed each of test.The tests performed were an endotoxin test, ph test, sterility test, and heavy metals and iron testing.All samples were found to be within specifications and there was no observed bacterial growth when plated and incubated.Bd was unable to identify a cause for the reported issue.Bd has placed controls in their manufacturing plants to prevent contamination such as weekly bioburden testing, an overkill sterilization process, biweekly environmental testing with the filling area, the solutions are filtered twice before filling, each sterilizer was thoroughly validated before use and a continuous online monitoring of the wfi water quality.There was no documentation of issues for the complaint of batch 716087n during this production run.There was no documentation of issues for the complaint of batch 723060n during this production run.Capa not required for this event.Complaints received for this product and condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGES
Common Device Name
HEPARIN FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key7519640
Report Number1911916-2018-00230
Device Sequence Number1862657
Product Code NZW
UDI-Device Identifier30382903064244
UDI-Public30382903064244
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K090680
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2019
Device Catalogue Number306424
Device Lot Number716087N
Initial Date Received by Manufacturer 05/01/2018
Supplement Date Received by Manufacturer05/01/2018
05/01/2018
Initial Report FDA Received Date05/16/2018
Supplement Report FDA Received Date05/31/2018
06/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-