PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER D; DXE Back to Search Results
Catalog Number CATD
Medical Device Problem Codes Kinked (1339); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Results: the catd was bent and ovalized starting approximately 30.0 cm from the hub to the distal tip.The catd was flattened from approximately 42.0 cm from the hub to the distal tip.The distal tip of the returned catd was twisted.Conclusions: evaluation of the returned catd revealed a flattened distal region.This damage may have been a result of the catheter tip shifting within its packaging during transit.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a thrombectomy procedure, the hospital staff noticed that the tip of an indigo system aspiration catheter d (catd) was kinked, flattened, and crushed upon removal from the packaging.The damage to the catd occurred prior to use.Therefore, the catd was not used in the procedure.The procedure was completed using a new catd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER D
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7520176
Report Number3005168196-2018-01004
Device Sequence Number19178745
Product Code DXE
UDI-Device Identifier00814548017624
UDI-Public00814548017624
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date02/22/2020
Device Catalogue NumberCATD
Device Lot NumberF74552
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2018
Initial Report FDA Received Date05/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-