Model Number ZXT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); Visual Disturbances (2140); No Code Available (3191)
Date of Event 12/13/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If explanted, give date: explant was planned.However not yet confirmed.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the surgeon was planning to explant a tecnis symfony toric multifocal intraocular lens (model zxt150, +13.0 diopter) from patient¿s right eye in secondary procedure because of patient experiencing cloudy vision, trouble with distance and near vision.At the time of this report, the lens has not yet been explanted.No further information provided.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7521063
Report Number9614546-2018-00475
Device Sequence Number19192078
Product Code POE
UDI-Device Identifier05050474605060
UDI-Public(01)05050474605060(17)220412
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/24/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date04/12/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/18/2018
Supplement Date Received by Manufacturer08/23/2018
Initial Report FDA Received Date05/16/2018
Supplement Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Date Report Sent to FDA01/01/2005
-
-