Model Number EM200-R
Medical Device Problem Codes Overheating of Device (1437); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.The device was tested and the temperature was measured to be 76.3°f.The rate of temperature rise was above threshold at 6°f/sec.The likely cause was identified as rear bearing worn and disintegrated, as a result the rotor shaft was not supported in the rear which caused the motor to vibrate due to rotor shaft being unbalanced.It was also noted that the motor was noisy, shedding debris were also found and the motor failed patient current leakage test.This type of failure is associated with (b)(4).The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of overheating.It was reported that there was no patient or staff impact.Repair request escalated to a product event based on reason for return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX® MOTOR LEGEND EHS STYLUS REFURB
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach street
fort worth TX 76137
MDR Report Key7521567
Report Number1625507-2018-00025
Device Sequence Number13747209
Product Code HBE
UDI-Device Identifier00613994422002
UDI-Public00613994422002
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K012456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM200-R
Device Catalogue NumberEM200-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2018
Initial Report FDA Received Date05/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-