| Catalog Number |
563514 |
| Medical Device Problem Codes |
Fracture (1260); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Pain (1994); Inadequate Osseointegration (2646); Not Applicable (3189)
|
| Date of Event |
04/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.No 510(k) number provided because this implant product code is sold internationally.It is sold in the us under a different product code.
|
| |
|
Event or Problem Description
|
|
Customer reported to bfarm: hip-tep right side on (b)(6) 1999.Osteolysis and pe wear.Revision of head and inlay in 2013.Now since months pain in the right hip.X-ray showed fractured stem.Punction of right hip: sterile: change of stem right hip on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
After review of the medical records on 3 jul 2018, it was discovered the patient underwent a revision surgery on (b)(6) 2013 due to loosening of the stem and sleeve as well as a greater trochanter fracture.A lesser trochanter fracture was discovered during the second revision on (b)(6) 2018.There is no evidence the stem and sleeve were removed during the first revision surgery on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: following inspection of the returned products and review of the information received, it was concluded that it was unlikely that a manufacturing defect was present.The complaint shall be closed with an undetermined conclusion, it shall be entered onto the complaints database and monitored through trend analysis.Should further information be received, then the complaint shall be reviewed and further investigation completed as required.Device history lot = > null device history batch = > null device history review = > null if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|