| Catalog Number |
8713040U |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2016018 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.Additional attempts to receive the device are being made.A follow-up report will be submitted when the results of the investigation are available.
|
| |
|
Event or Problem Description
|
|
As reported by user facility: syringe stopped infusing with additional volume to be infused.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number e2016018.(b)(4).Failure analysis and investigation results did not confirm the reported issue.Upon receipt the actual pump involved was visually and functionally inspected.Three (3) volumetrics of 100 ml/hr and 10 ml were performed and tested in spec.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number.The pump operated as intended and the reported failure could not be reproduced.Based on the results of the investigation, no conclusions can be made regarding the cause of the reported event.If additional information becomes available, a follow up report will be submitted.
|
| |
|