MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR) Back to Search Results
Model Number 8000.COM03
Medical Device Problem Code Device Operational Issue (2914)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/09/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has been investigated without conclusive result.During the retrospective review, and as per the complaint description, low supply of water is detected from the pump.Aspiration module error can lead to variation in pressure, hence considered reportable.The risk is low pressure or hypotony and is determined to be reportable.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre (b)(4)).Details of this activity were discussed with cdrh office of compliance (b)(4) during a tele-conference on 05/31/2017.No information reasonable suggests anyone was harmed.All available information has been disclosed.This complaint has now been closed.
 
Event or Problem Description
Error message, pum63 supply pressure switch on, supply pressure too low, water detected, aspiration module will be deactivated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR)
Common Device Name
EVA, COMBINED MACHINE INCLUDING LASER (SMA CONNECTOR)
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7522523
Report Number1222074-2018-00139
Device Sequence Number16610030
Product Code HQC
UDI-Device Identifier08717872019666
UDI-Public08717872019666
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 11/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number8000.COM03
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2015
Initial Report FDA Received Date05/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-