Pre-op diagnosis: primary osteoporosis type of fracture: compression fracture levels implanted: t12 it was reported that on (b)(6) 2018, intra-op, when performing the installation with adjusting the position of the left balloon before inflation, the tip of the balloon was stuck in the bone and could not be moved.Once the balloon was expanded and attempted to be removed, it did not move.The balloon was expanded with the right balloon at the same time, and was tried to be removed again.It could not be removed.The operation for removing it was repeated several times but it could not be removed.It could not be removed even removing the inner needle and performing the operation which could be assumed.The inner needle was removed, and the balloon could move by being twisted and turned back, but it could not be removed.According to the judgment of the physician, it would be removed with the preparation of the balloon ruptured.After deflating the balloon, it was twisted and the twisted part was blocked.Removing was performed in this state.The tube was cut halfway.The part which was cut was removed again with using kocher, but it could not be removed successfully.The needle external cylinder was inserted through the torn tube and removed.The tip of torn balloon was removed together with the external cylinder.It was removed successfully.It was considered that the tip part of balloon was chipped.There was a possibility that a slight amount of contrast media leaked.No patient complications were reported as a result of this event.According to the report, the balloon could not be removed.The physician also thought that there was a possibility that the balloon was stuck in the bone.A new one was opened to deal with it.
|