• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 04/20/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Udi not available for this system at time of filing.A medtronic representative went to the site to test the equipment.The navigation system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the surgeon monitor went black while in the verify instruments screen.They verified the cables were properly seated and verified the inputs were correct.After they separated the carts again, the case went on as planned.The issue only happened on the surgeon monitor, the staff cart was fine.There was no patient present when this issue was identified.
 
Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date provided.The monitor for the navigation system was returned to the manufacturer for analysis.The monitor was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7522600
MDR Text Key108622366
Report Number1723170-2018-02104
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-