| Model Number |
42843 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported there was leaking from the occluder of the delta valve when the surgeon tested the valve during the procedure.There was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.The valve met the requirements for siphon, and preimplantation testing.The valve did not meet requirements for reflux, pressure-flow and leak testing due to a tear in the top of the proximal occluder.It is unknown how or when this damage occurred.The instructions for use (ifu) that accompany the device caution that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, crushing or breaking of components.¿ the ifu also caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|