• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH RESOLUTION 360¿ CLIP; HEMOSTATIC METAL CLIP FOR THE GI TRACT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH RESOLUTION 360¿ CLIP; HEMOSTATIC METAL CLIP FOR THE GI TRACT Back to Search Results
Model Number M00521230
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/20/2018
Event Type  malfunction  
Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.The complainant was unable to report the lot number; therefore, the manufacture date and expiration date are unknown.It was reported the device was not used past its expiry date.(b)(6).(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a resolution 360 clip device was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, the case was completed successfully.The scope was then cleaned and used in an esophagogastroduodenoscopy (egd) procedure in the stomach of a new patient.The suction channel appeared to be clogged and multiple attempts were made to clear the obstruction.During manual decontamination, the clip from the first procedure was found lodged in the scope.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be okay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION 360¿ CLIP
Type of Device
HEMOSTATIC METAL CLIP FOR THE GI TRACT
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7522909
MDR Text Key108622023
Report Number3005099803-2018-01627
Device Sequence Number1
Product Code PKL
UDI-Device Identifier08714729875628
UDI-Public08714729875628
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00521230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-