Per service request the field service engineer (fse) went to the customer site and found that there were bubbles in the sweep flow line.The fse replaced the sweep flow coil (pn 6807701) and its fitting to resolve the issue.The repairs were verified as per the service procedures.Per act diff 2 ifu pn 4237495bc, "when plt backgrounds fail beckman coulter recommends for the following: check the system for bubbles, starting with the fluid reservoir and moving on to the sample and diluent pumps.Remove the bath shield and run a cycle, observing the fluid in the baths if necessary.Repair any leaks that have caused the bubbles, whether tubing, fitting, or component.' (b)(4).
|