Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were microscopically examined.The tip is damaged.There is shaft damage at the exit notch that is consistent with damage seen with the use of a guide wire.Functional testing was carried out by attaching an inflation device filled with water to the hub of the complaint device, and inflating the device to rated burst pressure.The device inflated and held pressure for 5 minutes without leaking.Inspection of the remainder of the device presented no other damage or irregularities.There was no evidence of any material or manufacturing deficiencies contributing to the confirmed tip damage and the reported difficulty.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|