| Model Number |
3505305M |
| Medical Device Problem Codes |
Off-Label Use (1494); Material Rupture (1546); Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
No Code Available (3191); Insufficient Information (4580)
|
| Date of Event |
04/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast reconstruction procedure with a mentor tissue expander 250cc device that deflated after implantation.The deflation was in the middle of the expander in the joint at the expander pad.As a result, the patient underwent explantation on (b)(6) 2018.The explanted device was not replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
On 06/12/2018, mentor completed the analysis on the suspect medical device.Upon receipt by mentor, the device returned without fluid.White material was observed within the device.No foreign material was observed on the shell surface.The product evaluation (pe) team discovered a rent measuring approximately 0.5 cm on the anterior aspect.In addition, a crease on the anterior aspect was noticed.No other anomalies were discovered.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, pe was precluded from further evaluating the device.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage.All the implants are 100% inspected before leaving the facility.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet; however, deflation trends for mentor saline-filled devices will continue to be monitored by quality assurance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: after clinical/secondary review of this file performed on 23-sep-2020, it was decided to add device code 1494 for off-label use to more accurately capture the reported event.The tissue expander device was implanted in the patient for longer than 6 months (implantation date: (b)(6) 2017, explantation date: (b)(6) 2018.) this is contraindicated in the ifu.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2020, mentor completed a second investigation on the suspect medical device to address the off-label usage of the device.Device evaluation summary: upon receipt by mentor, the device returned without fluid.White material was observed within the device.No foreign material was observed on the shell surface.The product evaluation (pe) team discovered a rent measuring approximately 0.5 cm on the anterior aspect.In addition, a crease on the anterior aspect was noticed.No other anomalies were discovered.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, pe was precluded from further evaluating the device.Based on the dates of implantation and explantation provided, it was noticed that the device was inside the patient for more than six months.Per mentor ifu, tissue expanders are intended for temporary subcutaneous or submuscular implantation and are not intended for use beyond six months.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage.All the implants are 100% inspected before leaving the facility.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet; however, deflation trends for mentor saline-filled devices will continue to be monitored by quality assurance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Upon review, mentor has determined that there is no evidence that this device was used in a breast tissue expansion procedure as initially reported.There is currently no information regarding the type of procedure.Manufacturer¿s reference number: (b)(4).Specified in h10 that the device is not confirmed to have been used in a breast tissue expansion application as initially reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: an incorrect patient code (3191 ¿ no code available) was used for this event.The correct code has been added.Manufacturer¿s reference number: (b)(6).H6 health effect - clinical code: e2401 "insufficient information".
|
| |
|
Additional Manufacturer Narrative
|
|
After additional review of this event by mentor's clinical safety and regulatory teams on 22-jul-2022, it has determined that the device was not used off-label.As a result, medical device problem code off-label use no longer applies.The device was implanted longer than six months, which is not in accordance with our instructions for use.Therefore h6 medical device problem code a23/use of device problem has been added.Manufacturer's reference number: (b)(4).H6 medical device problem code off-label use was replaced with a23/use of device problem.
|
| |
|