| Model Number |
MMT-751RNAS |
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that they experienced high blood glucose.The customer blood glucose level was over 490 mg/dl at the time of incident.Customer also stated that received alarmed for compromised force sensor system customer treated with manual injection.The drive support cap was flush.The product will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The pump alarmed compromised force sensor system during the basic occlusion test due to moisture damage on the force sensor resistor.Drive support cap was inspected and no anomaly was noted.
|
| |
|
Search Alerts/Recalls
|