| Catalog Number |
H1-LS |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pseudoaneurysm (2605)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Cine image analysis summary: the hawkone directional atherectomy device was not received for evaluation.No ancillary devices from the procedure were received.A single cine image was received for evaluation.The single cine image is of a contrast flow through a vessel.The radiopaque marker band of a support catheter and a guidewire are visible in the cine image.The vessel exhibits signs of a pseudo aneurysm.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician used a hawkone device for treatment of a 10mm very calcified non-cto lesion in the proximal right common iliac artery.The diameter was noted as 8-9mm.The vessel was reported to be moderately tortuous.The device was prepped as per ifu without any issue.The procedure was carried out successfully.It was reported, 2 weeks post procedure, in a follow-up visit, that a pseudo aneurysm was detected via ultrasound.Patient was brought back for a non-mdt covered stent.The patient is reported to be doing well.
|
| |
|