Batch review performed on 18 may 2018; lot 163797: (b)(4) items manufactured and released on 20 december 2016 expiration date: 2021-10-10; no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Other device involved: reverse shoulder system humeral reverse hc liner ø39/+3mm reference (k170452); lot 173396: (b)(4) items manufactured and released on 21 june 2017.Expiration date: 2022-05-29.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold with a similar reported event.Clinical evaluation performed by medical affairs (b)(4) on 18 may 2018; dislocation of a rsa required ablation of both components and reposition of the scapular glenosphere.The humeral component 145° was replaced with a 155° to increase stability.The reported metallosis is a consequence of the dislocation and not a defect of the devices.There no indications that these devices malfunctioned to cause the adverse event.
|