MMJ SA DE CV(USD) MALLINCKRODT; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 18870
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Extubate (2402)
Date of Event 05/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, while in use on a patient, sometime after intubation, air was supplied to the cuff but the device¿s cuff was not inflated.Recannulation of a replacement tube was required.Customer indicated that there was no patient harm.
 
Additional Manufacturer Narrative
Evaluation summary: one sample was received for evaluation, an inflation/deflation test was performed and it was observed the cuff deflated after few seconds, a visual inspection was performed, and it was observed the inflation line presents damage, the failure mode leak at inflation line is confirmed.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALLINCKRODT
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key7526027
Report Number2936999-2018-00349
Device Sequence Number1305184
Product Code BTR
UDI-Device Identifier20884521032528
UDI-Public20884521032528
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2021
Device Model Number18870
Device Catalogue Number18870
Device Lot Number16I0444JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Initial Date Received by Manufacturer 05/12/2018
Supplement Date Received by Manufacturer08/22/2018
Initial Report FDA Received Date05/18/2018
Supplement Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-