Model Number 802112
Medical Device Problem Codes Out-Of-Box Failure (2311); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service representative (fsr) confirmed the complaint.He found the pressure module was not fully seated.He pulled the module and reseated it.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that during the use of the device for a non-clinical activity, the pressure pod was causing a "?" to be displayed on the unit.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7526190
Report Number1828100-2018-00264
Device Sequence Number16546001
Product Code DTQ
UDI-Device Identifier00886799000724
UDI-Public(01)00886799000724(11)180227
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K172220
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number802112
Device Catalogue Number802112
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2018
Initial Report FDA Received Date05/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-