| Model Number |
802112 |
| Medical Device Problem Codes |
Out-Of-Box Failure (2311); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service representative (fsr) confirmed the complaint.He found the pressure module was not fully seated.He pulled the module and reseated it.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that during the use of the device for a non-clinical activity, the pressure pod was causing a "?" to be displayed on the unit.There was no patient involvement.
|
| |
|