Device evaluated by mfr: the device was retuned for analysis.The tip, distal shaft, collar and hypotube was microscopically and tactfully inspected.The tip is damaged.There are numerous shaft kinks.The stent system that was returned with the guidezilla ii has stent damage preventing the stent system to be inserted into the guidezilla ii.A 4.0mm test device was inserted into the collar of the guidezilla ii and advanced through the entire length of the device with no issues.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
|