Model Number 0606460
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device is pending return to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter has an alleged split that was noted during use, while patient was receiving iv therapy.There was no patient injury reported.
 
Event or Problem Description
It was reported that the catheter has an alleged split that was noted during use, while patient was receiving iv therapy.There was no patient injury reported.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was not returned for evaluation.Therefore, the investigation is inconclusive for the alleged catheter split as no objective evidence has been provided to confirm any alleged deficiency with the catheter.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date: 4/2022).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HICKMAN 10F TRIPLE-LUMEN CV CATHETER WITH SURECUFF TISSUE INGROWTH CUFF
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7526314
Report Number3006260740-2018-01011
Device Sequence Number16457278
Product Code LJS
UDI-Device Identifier00801741052002
UDI-Public(01)00801741052002
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0606460
Device Catalogue Number0606460
Device Lot NumberHUBR1642
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/27/2018
Supplement Date Received by Manufacturer01/22/2019
Initial Report FDA Received Date05/18/2018
Supplement Report FDA Received Date02/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-