| Model Number |
286735100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Mis poly driver tip broke when inserting iliac screw, surgeon tapped 7.0 and was inserting 7.5x80mm screw, he was able to keep the screw inside the patient and finish.The rod holder broke when passing the rod.Passing rod from iliac up to t10, skipped levels though patient consequence? : no.Is the information being submitted for this complaint all the details that are known/available regarding this event? : yes.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Visual examination at the macroscopic level revealed that the fracture was located at the rod holder¿s distal tip.Device was then sent for fracture analysis.The fracture analysis report reveals a static overloading failure.No material defects or other abnormalities were observed in this analysis.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the rod holder¿s distal tip becoming fractured cannot be positively determined.However, the fracture analysis report reveals a static overloading failure.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|