SMITH & NEPHEW, INC. SYN POR FEM COMP SZ 10; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC,CEMENTED OR UNCEMENTED
|
Back to Search Results |
|
| Model Number |
71306610 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
04/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, three months after implanting the cementless synergy stem, x-ray showed broken stem.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint device was not returned.The clinical/medical team concluded, the radiographic images confirm the reported finding.In the absence of pain, the explanted prosthesis, revision report and without information related to the patient's medical history, bone quality, fall/trauma history, a definitive root cause of the reported implant fracture cannot be determined.However, possible fatigue stress if micromotion was present or early weight bearing cannot be excluded as a contributing factor.The future impact to the patient beyond the revision cannot be concluded.Should any relevant clinical information be provided, this complaint would be re-evaluated.No further clinical/medical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: it was reported that three months after the procedure, the x-rays noted severe damage to the 3/4 stem without any pain.The affected synergy porous stem, used in treatment, was returned.A lab analysis was conducted and revealed what appears to be biological material embedded within the porous region.The material seems to be more prominent on the medial sides.There is practically no biologic material exhibited on the lateral side.This suggest a lack of proximal support and becomes important when examining the fracture sites.The polished femoral neck area exhibited scoring.It is not clear of the cause for this but nothing visually indicates this had any role in the stem fracture.Examination of the distal stem portion found scoring along the length.These score marks potentially came from contact with instrumentation during removal.This suggest the distal stem was fixed in the canal.The stem clearly failed in fatigue due to an overload.Sufficient details as to how the stem was over loaded have not been provided.Certainly the suggestions indicating the stem was not proximally supported would have played a role.It should also be stated the timing of 3 months implantation is short of the historical length of time for this type of issue.A review of the manufacturing records for the listed part did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident could not be corroborated.A medical analysis noted that the radiographic images confirm the reported finding.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint.
|
| |
|
Search Alerts/Recalls
|
|
|