| Model Number |
4968-35 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Fracture (1260); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 4968-60 lead, implanted (b)(6) 2016; 4968-35 lead, implanted (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a lead integrity alert for high sensing integrity counter (sic) on the right ventricular (rv) lead and a non-sustained high rate episode.The rv lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was later reported that the rv lead had a suspected fracture and the cardiac resynchronization therapy defibrillator (crt-d) was also suspected of having a noise issue.The rv lead was capped and the crt-d was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The additional information was previously submitted in regulatory report: 2649622-2018-10948.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was later reported that the high sic on the rv lead was due to oversensing of noise and the alerts continued to show on interrogation reports.
|
| |
|