| Catalog Number |
309592 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that after use a bd safety-lok¿ syringe malfunctioned as the safety device failed."the shield which should lock in place, pulled off the top of the needle after the injection" there was no report of exposure, injury or medical intervention needed.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident, therefore a root cause could not be determined.Dhr review for batch 7338970 (p/n 309578): manufacturing dates: 11/21/2017 to 11/26/2017.Batch quantity was (b)(4).Sleeve printing and sleeve assembly records were reviewed as part of this dhr review.All visual inspections were performed as per requirement with no quality notifications related to the complaint defect.Batch 7338970 was inspected and accepted based on meeting our inspection control plan and subsequently approved for shipment.The review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Root cause and capa not required as no defects were confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|