• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN SCW LWDG TRL S6 10D X 10P; PROSTHESIS,KNEEPATELLOFEMOROTIBIAL,SEMICONSTRAIND,CMNTD,POLYMR/METAL/POLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN SCW LWDG TRL S6 10D X 10P; PROSTHESIS,KNEEPATELLOFEMOROTIBIAL,SEMICONSTRAIND,CMNTD,POLYMR/METAL/POLYMR Back to Search Results
Catalog Number 71431754
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/23/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a revision surgery, the screw broke off wedge and is stuck in femoral component.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual examination of the returned legion screw wedge trial indicated that the screw top and bottom were missing, confirming the stated complaint.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A clinical analysis indicated that no clinical relevant documents were provided to conduct a thorough medical assessment.No medical assessment is warranted at this time.This complaint will be re-evaluated if more information becomes available.A review of complaint history did not reveal additional complaints for the listed part.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN SCW LWDG TRL S6 10D X 10P
Common Device Name
PROSTHESIS,KNEEPATELLOFEMOROTIBIAL,SEMICONSTRAIND,CMNTD,POLYMR/METAL/POLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key7528417
Report Number1020279-2018-00959
Device Sequence Number275897
Product Code JWH
UDI-Device Identifier03596010545961
UDI-Public03596010545961
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71431754
Device Lot Number10LSM0040
Initial Date Received by Manufacturer 04/23/2018
Supplement Date Received by Manufacturer04/23/2018
Initial Report FDA Received Date05/19/2018
Supplement Report FDA Received Date06/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-