SMITH & NEPHEW, INC. R3 48MM ID INTL COCR LINER 60MM; PRSTHSIS,HIP,SMI-CNSTRAIND,UNCMNTD,MTAL/POLYMR,NON-POROUS,CALICUM-PHSPHTE
|
Back to Search Results |
|
| Catalog Number |
71335860 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
04/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the left femoral stem broke inside the patient, so a surgical intervention was performed and the stem was removed.Also patient presented metallosis.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices were not returned.The clinical/medical team concluded, without the explanted components, pathology, or other supporting documentation, the reported metallosis cannot be confirmed nor can the root cause for the fractured stem be concluded.The patient impact beyond the left hip revision and associated pain cannot be determined.No further medical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches.Product was sterilized according to sterilization release documentation from quality control.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected complaint devices were returned and evaluated.A lab analysis conducted during this investigation noted that bone on-growth was observed on the porous coating on the shell; this indicates good fixation within the acetabulum.Scratches on the inner diameter of the shell were observed.The gouging on the shell¿s rim is likely due to removal.A footprint of the locking features of the shell was observed on the outer diameter of the liner.Bone on-growth was observed on the porous coated region of the sleeve; this indicates good proximal fixation of the femoral component.Damage and scratching on the sleeve and distal part of the stem are likely due to removal of the components.The stem fractured in the region proximal to the sleeve.Fretting was observed on the lateral side of the proximal stem.Radial lines were seen on the fracture surface of the proximal stem section.The crack most likely initiated in the lateral region of the stem and was due to fatigue.Scratches on the stem¿s neck were observed and were likely caused by removal.A circular wear zone was observed on the surface of the femoral head ; this is likely due to the articulation of the head within the liner in vivo.No defects in manufacturing of the device, material, or design were noted.A review of complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A clinical evaluation noted that without the pathology report or other supporting documentation, the reported metallosis cannot be confirmed.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|