| Model Number |
MA15-8NS |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacture date was unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure, the burr device snapped.It was reported that the device was being used with a bearing sleeve device.It was not reported if there were no delays in the surgical procedure or whether a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The exact date of this event is unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the reported condition that the cutter device had a broken tip was confirmed.The external features of the returned cutter were visually inspected and observed that the tip of the cutter was broken.The cutter was examined and photographed using 28x magnification.Based on the photographs taken the angle indicate that there was excessive force applied.The assignable root cause was determined to be due to excessive lateral or side to side force, which is user error.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|