• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNIV REV FEM BROACH REAMER; KNEE INSTRUMENT : BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNIV REV FEM BROACH REAMER; KNEE INSTRUMENT : BROACHES Back to Search Results
Catalog Number 961671
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the threads inside broach reamer are stripped.Nothing will thread down into it.No surgical delay.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIV REV FEM BROACH REAMER
Common Device Name
KNEE INSTRUMENT : BROACHES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7529209
Report Number1818910-2018-60038
Device Sequence Number1452891
Product Code HTO
UDI-Device Identifier10603295239772
UDI-Public10603295239772
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number961671
Device Lot NumberABB81354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Initial Date Received by Manufacturer 04/23/2018
Supplement Date Received by Manufacturer06/19/2018
Initial Report FDA Received Date05/21/2018
Supplement Report FDA Received Date06/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-