| Model Number |
6215 |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the balloon catheter was returned and analyzed.The mechanical operation of the balloon catheter was occluded.Visual analysis of the catheter indicated damage during use.The analyst noted the guidewire returned with the catheter stops backloading at 30.5 cm from the distal tip of the catheter due to occlusion of contrast.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during the implant procedure, the physician indicated there was difficulty introducing the guidewire into the balloon catheter.After introduction of the wire into the catheter, the guidewire became stuck at the distal end/tip of the balloon catheter.Introducing the wire by front loading allowed the wire to pass in and out of the tip with no noted issue.The catheter was removed.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|