• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/14 HOLE/301MM/LEFT; IMPLANT, FIXATION DEVICE, CONDYLAR PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/14 HOLE/301MM/LEFT; IMPLANT, FIXATION DEVICE, CONDYLAR PLATE Back to Search Results
Model Number 02.124.415
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional product codes: hrs, hwc.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent revision due to non-union of the distal femur fracture and the variable angle (va) locking compression (lcp) curved condylar plate was broken.The fracture had to be revised with plate and screws.There are broken screws.It was unknown if there was a surgical delay and as well as the patient outcome.No other additional information provided.Concomitant device reported: unknown screws (part# unknown, lot # unknown, quantity unknown) this complaint involves one (1) device.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Event description updated, concomitant devices reported are not concomitant and have been included as impacted product in this complaint.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent revision due to non-union of the distal femur fracture and the variable angle (va) locking compression (lcp) curved condylar plate and screws were broken.The fracture had to be revised with plate and screws.It was unknown if there was a surgical delay and as well as the patient outcome.The patient was implanted with a competitor's plate, original implant date is unknown, surgery was successfully completed.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part 02.124.415, lot 8949665: manufacturing site: mezzovico.Release to warehouse date: april 25, 2014.No non-conformance reports (ncrs) were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/14 HOLE/301MM/LEFT
Common Device Name
IMPLANT, FIXATION DEVICE, CONDYLAR PLATE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7529665
Report Number2939274-2018-52216
Device Sequence Number1422214
Product Code JDP
UDI-Device Identifier10886982042910
UDI-Public(01)10886982042910
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.124.415
Device Catalogue Number02.124.415
Device Lot Number8949665
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/30/2018
Supplement Date Received by Manufacturer05/17/2018
06/11/2018
Initial Report FDA Received Date05/21/2018
Supplement Report FDA Received Date05/22/2018
06/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-