• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL VALVE, BURR HOLE, 16MM, MEDIUM PRESSURE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL VALVE, BURR HOLE, 16MM, MEDIUM PRESSURE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42544
Medical Device Problem Codes Device Operates Differently Than Expected (2913); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the valve ¿did not function¿ during the procedure.The valve was never implanted.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that ¿no function¿ referred to no flow through the valve.
 
Additional Manufacturer Narrative
The returned valve was patent.The valve met the requirements for reflux, and leak testing.However, it did not meet the requirements for pressure-flow and preimplantation testing.There was debris noted in the interior and exterior of the device.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL VALVE, BURR HOLE, 16MM, MEDIUM PRESSURE
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7529708
Report Number2021898-2018-00258
Device Sequence Number1623830
Product Code JXG
UDI-Device Identifier00643169466913
UDI-Public00643169466913
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K831678
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number42544
Device Catalogue Number42544
Device Lot NumberE34899
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2018
Supplement Date Received by Manufacturer05/23/2018
06/15/2018
Initial Report FDA Received Date05/21/2018
Supplement Report FDA Received Date05/24/2018
06/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-