| Model Number |
SESR01 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant of the implantable pulse generator (ipg) was unsuccessful due to an unspecified set screw issue.The ipg was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.The returned device had foreign material in the connector.The returned device had a damaged connector.The returned device indicated a grommet issue.The returned device indicated a damaged grommet.The returned device indicated foreign material on set screw.The returned device indicated a set screw with a rounded socket.The returned device indicated a damaged set screw.If information is provided in the future, a supplemental report will be issued.
|
| |
|