| Catalog Number |
368659 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing of the bd vacutainer® push button blood collection set with pre-attached holder was cut.Found before use.No reports of serious injury or medical intervention noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for severed tubing with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for severed tubing with the incident lot was observed.Root cause description: based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.As a result, bd is reviewing specific areas in the manufacturing process where the cause of this issue may have originated.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|