• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  PINN MULTIHOLE W/GRIPTION 50MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  PINN MULTIHOLE W/GRIPTION 50MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121730050
Device Problem Disassembly (1168)
Patient Problems Pain (1994); Not Applicable (3189)
Event Date 04/23/2018
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Der states patient underwent a difficult primary total hip back in (b)(6) 2017 where a revision multihole pinnacle cup was implanted.The patient had a large cyst/defect in his acetabulum requiring large amounts of bone grafts and screws.A pinnacle altrx +4/10 degree was implanted along with summit stem and a 32+1 head.All components were very good choice for this patient.A few months later patient complained of painful hip revision surgery was required.Upon inspection intraoperatively it seemed the liner had not fully locked into pinnacle cup.A new liner was placed along with a new head.Fixation of liner was thoroughly inspected and locked properly into cup.Doi: (b)(6) 2017; dor: (b)(6) 2018; unknown hip.
 
Manufacturer Narrative
Product complaint # :(b)(4).Investigation summary :no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MULTIHOLE W/GRIPTION 50MM
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7530259
MDR Text Key108782856
Report Number1818910-2018-60106
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295009979
UDI-Public10603295009979
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071784
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Catalogue Number121730050
Device Lot NumberH81865
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2018
Date Device Manufactured05/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-