• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 CERAMAX CERAMIC INSERT 36 X 56; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 CERAMAX CERAMIC INSERT 36 X 56; PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS Back to Search Results
Catalog Number 121887656
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); Not Applicable (3189)
Date of Event 04/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Der states that during a routine left tha replacement, a 56mm pinnacle cup was implanted.No screws were placed in the cup.A ceramax 36x56 liner was inserted after fixation of the pinnacle cup.The surgeon checked to make sure the liner was properly inserted prior to impaction.Upon impaction the liner fractured in several places.The broken liner was removed, and a new 36x56 ceramax liner was inserted.The 2nd liner was impacted and seated correctly.Please check returned fractured liner to make sure it is within proper specifications.Doe: (b)(6) 2018.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary : the device was returned and failure mode confirmed.The insert had fractured.A review of complaint databases and manufacturing records did not identify any anomalies.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERAMAX CERAMIC INSERT 36 X 56
Common Device Name
PINNACLE HIP SYSTEM : HIP CERAMIC ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7530526
Report Number1818910-2018-60131
Device Sequence Number1452881
Product Code MRA
UDI-Device Identifier10603295012535
UDI-Public10603295012535
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121887656
Device Lot Number8482507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2019
Initial Date Received by Manufacturer 04/23/2018
Supplement Date Received by Manufacturer01/08/2019
03/25/2019
Initial Report FDA Received Date05/21/2018
Supplement Report FDA Received Date01/28/2019
03/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age25 YR
-
-