| Model Number |
LS1520 |
| Medical Device Problem Codes |
Retraction Problem (1536); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a hysterectomy procedure, tissue was sealed finely as usual, but then the knife did not return and the device could not be used anymore.A new device was opened to continue.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The reported conditions were not confirmed.The jaws opened and closed properly when the latch was retracted and released.The knife blade advanced and retracted smoothly along the knife track.The knife cut of the device was tested on a silicone test strip with acceptable results.The knife blade was also inspected under microscope and no issue was found.The investigation found the device to function normally and within specifications.The instruction for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|