Catalog Number B2060-200
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number -(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspection was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not identify a lot specific product related issue.It is likely that the reported balloon rupture occurred due to interaction with the lesion site.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during a procedure of the moderately calcified, chronic totally occluded, left superficial femoral artery (sfa) the 6.0 x 200 mm armada balloon dilatation catheter (bdc) was inflated below or at rated burst pressure (rbp) when the balloon ruptured.The device was removed and replaced without issue.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7530821
Report Number2024168-2018-03783
Device Sequence Number1652109
Product Code LIT
UDI-Device Identifier08717648155321
UDI-Public08717648155321
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2020
Device Catalogue NumberB2060-200
Device Lot Number61007G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2018
Initial Report FDA Received Date05/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight77
-
-