Catalog Number 21-7036-01
Medical Device Problem Code Insufficient Flow or Under Infusion (2182)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
Information was received indicating that the patient using this administration set had a peripherally inserted central catheter that was being "sluggish" over the course of two days.It was noted that 200 ml of 600 ml of fluids were left in their iv bag when the pump read that only 10.5 ml were left in the bag.The pump programming was checked and confirmed to be set correctly.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
One infusion administration set was returned for evaluation.The sample was set for accuracy testing using a pump solis vip, and no discrepancies were detected.While no definitive root cause to the reported issue could be determined, this investigation revealed no intrinsic evidence to suggest a root cause related to manufacturing.No corrective actions are required since the complaint was not confirmed and the root cause could not be attributed to the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD® CADD® ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key7531227
Report Number3012307300-2018-01614
Device Sequence Number10808986
Product Code FPA
UDI-Device Identifier10610586020186
UDI-Public10610586020186
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K933390
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number21-7036-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2018
Supplement Date Received by Manufacturer08/30/2018
Initial Report FDA Received Date05/21/2018
Supplement Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-