• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM LOCKING SCREW SLF-TPNG W/STARDRIVE(TM) RECESS 16MM; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.5MM LOCKING SCREW SLF-TPNG W/STARDRIVE(TM) RECESS 16MM; NAIL, FIXATION, BONE Back to Search Results
Model Number 212.104
Medical Device Problem Code Failure To Adhere Or Bond (1031)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient age/date of birth and weight are unknown.(b)(4).Due to intra-operative issues, the device was not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018 during an unknown procedure, an 8-hole 1/3 tubular plate was placed on the patient's distal fibula.After drilling through the drill guide, the locking screw would not lock into the plate.The surgeon tried another screw and the screw locked in.The patient outcome was stable and the procedure was successfully completed.There was a four (4) minute surgical delay and was no patient consequence reported.No other additional information available.Concomitant device: unknown drill guide (part unknown, lot unknown, quantity 1).This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
A product investigation was completed: visual inspection performed showed no wear or damage or abnormalities to the returned screw when viewed under 10x magnification.A functional test in effort to replicate the reported complaint condition was not able to be performed because the plate involved in the event was not returned.The reported complaint condition of the screw not locking to the plate was not able to be confirmed or replicated.A review of the device history records was unable to be performed since the lot number was unknown.Relevant drawings for the family of 3.5mm locking screws was reviewed during this investigation.No product design issues or discrepancies were observed.The peak major diameter (most proximal) of the returned locking screw head was measured and was within specification.The major thread diameter of the returned locking screw shaft was measured and was within specification.A definitive root cause for the reported complaint condition could not be determined based on the provided information.It is possible that the screw was not aligned perpendicular to the plate hole thread during attempted insertion.This complaint was not able to be confirmed and no product design issues or manufacturing discrepancies that would contribute to the reported complaint condition were identified during this investigation.No new, unique or different patient harms were identified as a result of this evaluation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Concomitant device: unknown drill guide (part: unknown, lot: unknown, quantity: 1), locking compression plate (lcp) cloverleaf 3.5mm 3-hole sst (part: 241.831, lot: unknown, quantity: 1).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5MM LOCKING SCREW SLF-TPNG W/STARDRIVE(TM) RECESS 16MM
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7531330
Report Number2939274-2018-52248
Device Sequence Number122422
Product Code JDS
UDI-Device Identifier10886982151780
UDI-Public(01)10886982151780
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K072095
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number212.104
Device Catalogue Number212.104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Initial Date Received by Manufacturer 05/01/2018
Supplement Date Received by Manufacturer07/06/2018
08/02/2018
Initial Report FDA Received Date05/21/2018
Supplement Report FDA Received Date07/26/2018
08/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-