• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE COYOTE¿; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE COYOTE¿; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number H74939186202210
Device Problem Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/25/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device evaluated by mfr: the device was not received for analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that catheter entrapment occurred.The 90% stenosed target lesion was located in the mildly tortuous and moderately calcified vessel below the knee.After a non-bsc guide wire was passed through the lesion, a 2.0mm x 220mm x 150cm coyote¿ balloon catheter was advanced for pre-dilatation an reached the area before the lesion and would not advance any further.The device was stuck with the guide wire.It was a long procedure and thrombus adhesion to the wire can be suspected.Both devices were removed together and the procedure was completed with a different device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE¿
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7531426
MDR Text Key108856695
Report Number2134265-2018-04239
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
PMA/PMN Number
K111295
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Model NumberH74939186202210
Device Catalogue Number39186-20221
Device Lot Number19405936
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/25/2018
Initial Date FDA Received05/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
COMMAND GUIDEWIRE
-
-