• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COSMAN MEDICAL NITINOL TC ELECTRODE; PROBE, RADIOFREQUENCY LESION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COSMAN MEDICAL NITINOL TC ELECTRODE; PROBE, RADIOFREQUENCY LESION Back to Search Results
Model Number TCN-10
Medical Device Problem Codes Material Discolored (1170); Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Expiration date: na.A total of (5) electrodes were analyzed: tcn-10, lot 011116, quantity (1), nitinol tc electrode, 100 mm the device analysis performed on the (1) electrode revealed discolored epoxy and a chip out.The color of newly cured epoxy is amber, light brown.If the epoxy is subjected to too many autoclave cycles it becomes brittle and discolored.This failure mechanism is unrelated to the reported complaint and is due to external mechanical force and thermal stress associated with multiple autoclave cycles.Tcn-15 lot 051016, quantity (3) nitinol tc electrode, 150 mm the device analysis performed on the (3) electrodes revealed discolored epoxy and a chip out.The color of newly cured epoxy is amber, light brown.If the epoxy is subjected to too many autoclave cycles it becomes brittle and discolored.This failure mechanism is unrelated to the reported complaint and is due to external mechanical force and thermal stress associated with multiple autoclave cycles.Csk-tc10, lot n184, quantity (1) csk tc electrode, 10cm the device analysis performed on the (1) electrode was found to be completely separated from the hub due to excessive external mechanical force.The shaft was not returned for analysis.The functional test could not be performed due to the separated shaft.
 
Event or Problem Description
The device analysis performed on the five electrodes revealed that four of them had discolored epoxy and a chip out.The color of newly cured epoxy is amber, light brown.If the epoxy is subjected to too many autoclave cycles it becomes brittle and discolored.This failure mechanism is unrelated to the reported complaint and is due to external mechanical force and thermal stress associated with multiple autoclave cycles.The device analysis performed on the fifth electrode was found to be completely separated from the hub due to excessive external mechanical force.The shaft was not returned for analysis.The functional test could not be performed due to the separated shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITINOL TC ELECTRODE
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key7531900
Report Number3006630150-2018-01781
Device Sequence Number156962
Product Code GXI
UDI-Device Identifier00813250013054
UDI-Public00813250013054
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K50084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model NumberTCN-10
Device Lot Number011116
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2018
Initial Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-