| Catalog Number |
71179215 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that towards the end of the surgery it looked like the devices needed more force and the surgeon kept turning the drivers until it snapped off.Four drivers were broken in the same way.There was minimal delay to remove screw driver fragments from patient.Doctor verified with x rays that the devices were removed.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated devices were returned and evaluated.A visual inspection of the returned conquest fn screwdriver indicated that a portion of the hex features have fractured off.The fractured off portions of the devices were not returned.After evaluating the device and reviewing the dhf it is unclear the exact cause of failure.The modification of the subject device was evaluated and was found to have an acceptable functionality.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.A clinical analysis indicated no clinical information has been provided for inclusion in this investigation.All four drivers were reported broken.Based on the quality review of the dhf the exact cause of the failure is unclear.The future impact to the patient cannot be determined.No further clinical/medical assessment is warranted.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|