Catalog Number 47369802
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that packages of bd¿ pen needle sterile single use were found with defective covering, missing or punctured/tore breaching the sterile barrier.There was no report of exposure, serious injury or medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: on (b)(6) 2018, the (b)(4) manufacturing plant received the following complaint samples: twenty (20) pen needles from batch# 7152646: defect(s) reported: during 100% inspection, 9 critical defects were identified: quantity: 8; defect: missing, damaged or defective label affecting sterility (3 punctured/teared; 5 not sealed) quantity: 1; defect: missing, damaged, or defective cover affecting sterility.Additionally, 13 major defects of interest that appear to be of similar root cause as the critical defects were found during 100% inspection: damaged, deformed, or defective cover and/or label not affecting sterility (11 cuts, tears to label and/or slight damage to cover where label damage occurs; 2 bottom hit to cover).A device history record review for catalog# 47369802, batch# 7152646 (23g etw x 7mm pen needle us).This batch was manufactured between 18aug2017 to 28aug2017.There were zero (0) notifications for any defects noted during manufacturing of this batch.Investigation summary: damage noted to the exterior of the cover and to the tear drop label, as noted in the sample photos above, was determined to be attributable to a previously unidentified station.At station 50, a paper support ledge is utilized to support the flow of label material.An unevenness between this support and the adjoining plate was found to be a contributing factor to damage found along the rim and/or side of the covers.Damage to the tear drop label was also noted while observing the production line running during evaluation in (b)(6) 2017.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ PEN NEEDLE STERILE SINGLE USE
Common Device Name
INSULIN PEN NEEDLE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key7532210
Report Number1920898-2018-00332
Device Sequence Number275882
Product Code FMI
UDI-Device Identifier00382904736989
UDI-Public00382904736989
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K110105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2022
Device Catalogue Number47369802
Device Lot Number7152646
Initial Date Received by Manufacturer 04/27/2018
Supplement Date Received by Manufacturer04/27/2018
Initial Report FDA Received Date05/22/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-