| Model Number |
R SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self test.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed and attributed to a faulty test port.This is a malfunction which only impacts the self-test portion with the defibrillator.Once the multi-function connection is removed from the test port and attached to a set of electrode pads or external paddles the device performs to specification.This type of malfunction does not pose any clinical impact and does not meet our guidelines for reportability.Analysis of reports of this type has not identified an increase in trend.
|
| |
|