ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Date/Time-Related Software Problem (2582)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to animas.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging that the time and date reset to default randomly.There was no indication that the device caused or contributed to an adverse event.This complaint is being reported because of a time change during use, without user intervention, may result in the user running the incorrect basal segment leading to over or under delivery.
 
Additional Manufacturer Narrative
The device has been returned and evaluated by product analysis on 30-may-2018 with the following findings: a review of the black box did not show a time and date reset to default setting.The pump was tested and no time and date reset to default setting was duplicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7533155
Report Number2531779-2018-09420
Device Sequence Number19210656
Product Code OYC
UDI-Device Identifier10840406100280
UDI-Public10840406100280
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 05/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2018
Was the Report Sent to FDA? Yes
Device Age16 MO
Initial Date Received by Manufacturer 05/01/2018
Supplement Date Received by Manufacturer05/01/2018
Initial Report FDA Received Date05/22/2018
Supplement Report FDA Received Date06/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age15 YR
Patient Weight75
-
-